Bite marks on TEE probes.
The shaft of a TEE probe lies between the patient's teeth. Even with a bite guard, cuts arise in the bending rubber and the shaft gets crushed. Fluid enters through the cuts and corrosion begins in the articulation mechanics.
How to spot it
Visual inspection of the whole shaft after every examination, especially in the area of the teeth. Plus the leak test — on TEE probes it is not optional. A probe with damaged bending rubber must not be used again.
What we do
We repair the bending rubber, treat the articulation mechanics against corrosion and then test leakage current and tightness. With advanced damage, repair of the distal end follows, or a complete overhaul as Refurbished by MOONS.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.