Lens crack and lens wear.
A crack in the lens is more than an image problem: it is a cross-contamination risk, because material collects in the gap that wipe disinfection can no longer reach. Alcohol-based agents dry the lens out further and accelerate cracking.
How to spot it
Visual inspection under good light and, if possible, with a magnifier: fine hairline cracks, rough patches, material loss at the edges. A probe with a visible crack on the patient-contact surface belongs out of service, even if the image is still usable.
What we do
We renew the coating completely and then test leakage current and tightness so the probe can safely be used on patients again. The result is in the test report that goes back with the probe.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.
Probe types affected
- Standard & matrix probes
- Wireless probes
- 3D/4D & matrix 3D/4D
- TEE & matrix TEE probes
Send in a probe with this fault
Five questions in the repair advisor, and at the end you get a completed enquiry form with an initial assessment.