100 questions on probe repair — answered.
Everything hospitals, practices and biomedical engineering departments want to know before placing an order. Every answer ends with the next step — usually one click into the repair advisor, which classifies your case straight away.
Go to the repair advisorCost, quotation, decision
Why there is no price list, how binding the quotation is and when a repair is worth it.
Why are there no prices on this site?
Because an honest price is only possible after the analysis. Two probes of the same model can mean completely different work — one has a worn strain relief, the other has damaged crystals. A price list online would either be set too high or lead to add-on charges later. We want neither.
Next stepSend an enquiry — answer within one working day→Is there a charge for the analysis if I decline the quotation?
Ask us about this before you send the probe — we settle it in writing beforehand so there are no surprises. The repair itself never starts until you have approved it.
Next stepSend an enquiry — answer within one working day→How binding is the quotation?
It is based on the actual measurement, not on a remote estimate. If additional damage shows up once the probe is open, we come back to you before doing any further work. You never receive an invoice for work you did not approve.
Next stepSend an enquiry — answer within one working day→Is it worth repairing an older probe at all?
Often it is exactly then. Replacement probes for older systems are hard to source, and where they exist the price is out of all proportion. As long as the transducer is intact, almost everything can be restored. If it is not worth it, we say so — before the repair, not after.
Next stepClassify your case in the repair advisor→How much do I save compared with a new probe?
Usually up to 70 percent. How much in your case depends on the repair level: reconditioning a connector costs a fraction of a full overhaul, and even that is well below the price of a new probe.
Next stepThe repair levels in detail→Can we agree a framework contract or fixed terms?
Talk to us. For hospitals with many probes and recurring demand this can be arranged sensibly. Call +43 316 22 81 10 or note it on the enquiry protocol.
Next stepSend an enquiry — answer within one working day→Can we send several probes at once?
Yes. Enclose a separate enquiry protocol with each probe so the allocation stays unambiguous — every probe gets its own reference number, its own quotation and its own test report.
Next stepEnquiry protocol as a PDF→What if there are several faults at the same time?
Then a higher level usually applies, because the work is combined. That is cheaper than opening the same probe twice. The quotation states which combination is needed.
Next stepThe repair levels in detail→What does repairing an ultrasound probe cost?
Every repair is calculated individually, because the effort depends on the probe model and its condition. After the analysis in Graz you receive a written quotation. Compared with buying new you usually save up to 70 percent. We only begin work once you approve the quotation.
Next stepSend an enquiry — answer within one working day→How long does a probe repair take?
On arrival the probe goes through our test chain, after which you get the quotation — we reply to your enquiry within one working day. The repair itself depends on the scope and on the availability of spare parts. If the parts are there, a 24h express is possible. For the meantime we provide a loan probe subject to availability.
Next stepThe repair levels in detail→Process, turnaround, shipping
From enquiry to return: collection, packing, express and checking the status.
How does a repair run from start to finish?
Enquiry through the protocol, collection across Europe on request, analysis in Graz along our test chain, written quotation, your approval, repair, final inspection, return shipment with the test report. Nothing happens between step four and step five without your approval.
Next stepSend an enquiry — answer within one working day→How quickly do I get an answer to my enquiry?
Within one working day. The more completely the enquiry protocol is filled in, the more likely that answer is already an assessment rather than a follow-up question.
Next stepSend an enquiry — answer within one working day→What exactly does 24h express mean?
That the repair is given priority after your approval, provided the parts needed are in stock. If parts have to be sourced, express does not help — then it depends on the delivery time. We tell you which case applies in the quotation.
Next stepSend an enquiry — answer within one working day→Do you collect the probe?
On request we arrange transport across Europe. Tick collection on the enquiry protocol and we will come back to you with a date.
Next stepSend an enquiry — answer within one working day→How should I pack the probe?
Shock-protected, with the connector and the transducer head padded separately, the cable coiled loosely — not wound tightly around the handle. Use the original case if you have it. Enclose the printed enquiry protocol.
Next stepEnquiry protocol as a PDF→Should I send the transport case with it?
If you have it, please do — it is the best protection in both directions. Note it on the protocol and it comes back with the probe.
Next stepEnquiry protocol as a PDF→Can I bring the probe in myself?
Yes, by arrangement. The workshop is in Graz, Weinholdstraße 33. Call ahead so someone from the technical team is free.
Next stepSend an enquiry — answer within one working day→Is the transport insured?
When we arrange collection, transport runs through our service providers. Agree the insured value with us beforehand, especially for matrix, TEE and endoscope probes.
Next stepSend an enquiry — answer within one working day→Do you repair probes sent from outside Europe?
We arrange collection across Europe. Probes from other regions can be sent to us — please clarify customs and the shipping route with us beforehand so the consignment does not get stuck.
Next stepSend an enquiry — answer within one working day→How do I check the status of my probe?
Through the reference number from the repair advisor, in the format MR-YYYYMMDD-XXXX. Quote it with any query and the case is identified immediately.
Next stepClassify your case in the repair advisor→May I have my probe repaired outside the manufacturer?
Yes. Maintenance of medical devices is not tied to the manufacturer; any business meeting the professional requirements may be commissioned. Whoever assesses the probe professionally must be free from instructions in doing so. Manufacturer specifications must be taken into account but do not establish a service monopoly. Whether a running device warranty or a full-service contract contains its own rules is stated in that contract — best checked beforehand.
Next stepSend an enquiry — answer within one working day→Which manufacturers do you repair?
We work brand-independently: GE Healthcare, Philips including the legacy stock of ATL and HP, Siemens Healthineers and Acuson, Canon and Toshiba, Esaote, Fujifilm with Hitachi, Aloka and Sonosite, Mindray, Samsung with Medison and SonoAce, SonoScape, Alpinion, BK Medical, Olympus for endoscopes and Butterfly. In total 915 model designations are on file. If your probe is not among them, ask anyway.
Next stepSend an enquiry — answer within one working day→Repair levels
What the five levels cover, how the classification works and what Refurbished by MOONS means.
How is the repair level determined?
From the findings of the test chain. The repair advisor on this site derives an initial assessment from your answers — the level only becomes binding after the measurement in our workshop.
Next stepClassify your case in the repair advisor→What does level 1 cover?
Cosmetic repairs, partial cable repairs and cleaning the connector pins. It is the smallest intervention and makes sense for probes that work but show wear or tarnished contacts.
Next stepThe repair levels in detail→What does level 2 cover?
Housing, strain relief, connector, and repair of the pins or renewal of the locking pin. Typical after drops, with a worn latch or a damaged cable exit.
Next stepThe repair levels in detail→What does level 3 cover?
The coating including partial reconditioning. This is the standard route for a delaminated or worn lens — the most common finding of all.
Next stepDelaminated lens: how to spot it and what to do→What does level 4 cover?
Package repairs: the coating combined with strain relief, housing or cable reconditioning, plus element leads. Sensible when several assemblies are affected and a second opening should be avoided.
Next stepThe repair levels in detail→What does level 5 cover?
The full overhaul including crystal replacement, with twelve months of warranty. It is the largest intervention and comes into question when the transducer itself is affected.
Next stepThe repair levels in detail→Why do 3D/4D probes have only four levels?
Because they are built differently: instead of a crystal replacement, the top step is the complete overhaul of the mechanics, the element leads and the electronics. That highest level is called Refurbished by MOONS.
Next stepClassify a 3D/4D probe in the repair advisor→What does Refurbished by MOONS mean?
The complete overhaul as the highest level for 3D/4D probes, TEE probes and endoscopes. For those types the upper levels carry this name instead of the term full overhaul.
Next stepThe repair levels in detail→Can I specify a level myself?
You can express a preference, but it rarely helps. The level needed follows from the measurement. We recommend the smallest level that makes the probe safe and diagnostically usable again.
Next stepClassify your case in the repair advisor→Is a probe as good as new after the repair?
After a full overhaul with crystal replacement it comes very close, and we give twelve months of warranty on it. At smaller levels exactly what was defective is repaired — the test report shows the condition in which it goes back.
Next stepThe repair levels in detail→Do I get a loan probe during the repair?
Subject to availability we provide a loan probe for the duration of the repair so your department keeps working. State the need right in the enquiry form and we reserve it together with the scheduling.
Next stepThe repair levels in detail→Probe types
Standard, 3D/4D, matrix, TEE, intraoperative, endoscopes — and what to watch for in each design.
Which probe types do you repair?
Standard probes (linear, convex, sector), 3D/4D volume probes, matrix probes, TEE probes, intraoperative and special probes, and endoscopes. Each type has its own level logic because the construction differs fundamentally.
Next stepClassify your case in the repair advisor→Do you repair linear probes?
Yes, linear probes belong to the standard probes and are the most frequent case in the workshop. Typical findings are delaminated lenses, cable breaks at the handle exit and damaged connector pins.
Next stepClassify a standard probe in the repair advisor→Do you repair convex probes?
Yes. Abdominal and convex probes run through the standard levels. The most common finding is a worn or lifting lens, often together with drop damage to the housing.
Next stepClassify a standard probe in the repair advisor→Do you repair sector and phased-array probes?
Yes. Cardiology sector probes run through the standard levels, up to the full overhaul with crystal replacement.
Next stepClassify a standard probe in the repair advisor→Do you repair 3D/4D volume probes?
Yes. Here mechanics come on top of the electronics: the transducer is moved inside the housing, and that drive is a typical source of failure. The highest level is the complete overhaul, Refurbished by MOONS.
Next stepClassify a 3D/4D probe in the repair advisor→Do you repair matrix probes?
Yes. Matrix probes have far more channels than classic arrays, which makes diagnosis more involved — element failures often only show up in the channel measurement. Pack these probes with particular care.
Next stepClassify a standard probe in the repair advisor→Do you repair TEE probes?
Yes. TEE probes are the most sensitive type because the shaft is used invasively. That is why a leakage test is always part of the test chain, and the highest level is called Refurbished by MOONS.
Next stepClassify a TEE probe in the repair advisor→Why is a leakage test mandatory on TEE probes?
Because an undetected leak in the shaft lets disinfectant and body fluid into the electronics — and the patient is examined with a probe that is contaminated inside. The test is not an extra, it is part of every TEE job.
Next stepClassify a TEE probe in the repair advisor→Do you repair endoscopes?
Yes, endoscopes are a field of their own with their own level structure, whose highest step is the complete overhaul. Describe what you notice on the enquiry protocol — image loss, leak tightness, angulation or mechanics.
Next stepClassify a flexible endoscope in the repair advisor→Do you repair intraoperative and special probes?
Yes. Hockey-stick, laparoscopy, biopsy and other special shapes fall into this group. Give us the exact model designation, because the designs differ widely.
Next stepClassify a standard probe in the repair advisor→Do you repair endocavity probes?
Yes. Transvaginal and transrectal probes run through the standard levels. When shipping, make sure the probe is disinfected and the hygiene declaration is enclosed.
Next stepClassify a standard probe in the repair advisor→Do you also repair batteries and charging units?
Yes, and it comes from our origins: AccuPower has been manufacturing and assembling battery packs since 1998. Give us the device and the battery designation and we will tell you whether a cell replacement or a new build makes more sense.
Next stepClassify a wireless probe in the repair advisor→Do you repair cables and connectors on their own?
Yes. Cable breaks, strain relief, connector housings, pins and the locking pin are classic level 1 and level 2 jobs — frequently without touching the transducer at all.
Next stepCable break and damaged strain relief: how to spot it and what to do→What about handheld and CMUT probes?
Send us the exact type designation and the fault. With newer semiconductor transducers we assess case by case what can sensibly be repaired — a blanket promise would not be serious.
Next stepSend an enquiry — answer within one working day→Fault patterns and diagnosis
How to recognise the defect, what to do immediately and how we measure.
Which fault patterns do you cover?
The eight most common are: delaminated lens, cable break, drop damage to the housing, element failure with stripes in the image, connector damage, fluid ingress, a leaking TEE shaft and failure of the 3D/4D mechanics. Each has its own page with the signs to look for.
Next stepAll fault patterns at a glance→How do I recognise a delaminated lens?
By bubbles, ripples or soft spots on the contact surface, often with a poorer near field or a dull area in the image. Stop using the probe — an open lens is a hygiene risk.
Next stepDelaminated lens: how to spot it and what to do→How do I recognise a cable break?
By the image flickering when the cable is moved, single lines dropping out, or the system briefly failing to detect the probe. The break is frequently right at the handle or connector exit, where the cable is bent most.
Next stepCable break and damaged strain relief: how to spot it and what to do→What do dark stripes in the image mean?
Usually an element failure: individual crystals or their leads no longer transmit. The stripes stay at the same position in the image even when you rotate the probe — that is what distinguishes them from artefacts.
Next stepImage dropout and shadowing: how to spot it and what to do→What should I do after a drop if the image still looks fine?
Have it checked anyway. A drop can leave hairline cracks in the housing or micro-fractures at the transducer that only cause failure weeks later — and a crack makes the probe hygienically vulnerable long before the image deteriorates.
Next stepClassify a standard probe in the repair advisor→The system no longer detects the probe. What causes that?
Frequently the connector: bent or corroded pins, a worn locking pin, a damaged connector housing. First test the same port with another probe so it is clear whether the fault is in the probe or in the system.
Next stepConnector and pin faults: how to spot it and what to do→The probe has been immersed in disinfectant. What now?
Take it out of service immediately, do not plug it in again and do not try to blow-dry it. Fluid inside starts corrosion that continues with every hour. Report the probe quickly — the sooner it is open, the more can be saved.
Next stepFluid ingress: how to spot it and what to do→How do I recognise a leaking TEE shaft?
By discolouration, soft or swollen areas on the shaft, moisture after reprocessing, or a conspicuous result in your in-house leak test. On any suspicion: quarantine the probe and have it tested.
Next stepBite marks on TEE probes: how to spot it and what to do→The 3D/4D probe hums or no longer builds a volume.
That points to the mechanical drive — motor, gearing or bearings. Do not keep operating the probe, because a blocked drive causes consequential damage to the mechanics.
Next stepMechanics in 3D/4D probes: how to spot it and what to do→Can I keep using the probe until the repair?
Not with a damaged housing, an open lens or suspected fluid ingress. There it is not about image quality but about patient safety and electrical safety. With purely cosmetic marks that is a different question — describe the finding to us.
Next stepSend an enquiry — answer within one working day→How do you test a probe? What is the test chain?
Four steps: incoming inspection and documentation of the condition, electrical measurement of all channels, image test on an ultrasound system, visual and leak-tightness inspection. For TEE probes and endoscopes the leak test is mandatory on top.
Next stepClassify your case in the repair advisor→Do you also find faults we have not noticed?
That is the normal case. The channel measurement shows weak elements before they become visible in the image, and the visual inspection finds hairline cracks nobody sees in use. Everything we find goes into the quotation — not onto the invoice without your approval.
Next stepClassify your case in the repair advisor→What if you find no fault?
Then we say so and you receive the test report with the measured values. Not infrequently the fault turns out to lie with the system, the port or the settings — and documented proof that the probe is sound has a value of its own.
Next stepSend an enquiry — answer within one working day→How precisely should I describe the fault?
As concretely as possible: since when, in which mode, constantly or intermittently, after a drop or after reprocessing, at which point in the image. A screenshot of the faulty image helps enormously. The better the description, the faster the assessment.
Next stepClassify your case in the repair advisor→Should I send a photo or a screenshot?
Yes, please — a photo of the affected area and a screenshot of the image fault. Both can be attached to the contact form and often save a follow-up question.
Next stepSend an enquiry — answer within one working day→What is the repair advisor on this site?
A short path of questions: probe type, manufacturer, model, fault. At the end there is a likely classification and an enquiry protocol with a reference number. The classification is guidance, not a commitment — the measurement is what counts.
Next stepClassify your case in the repair advisor→What do the five repair levels mean?
Level 1 is cosmetic work and partial cable repairs. Level 2 covers housing, strain relief, connector and pins. Level 3 is the coating with partial refurbishment. Level 4 are package repairs combining several areas. Level 5 is the full overhaul including crystal replacement with twelve months warranty. 3D/4D probes have four levels; on TEE probes and endoscopes the highest stage is called Refurbished by MOONS.
Next stepAll fault patterns at a glance→Manufacturers and models
Which brands we work on, where to find the model designation and how the manufacturer warranty is affected.
Which manufacturers do you work on?
GE, Philips, Siemens Healthineers, Canon (Toshiba), Esaote, Fujifilm (Sonosite/Hitachi), Mindray, Samsung Medison, SonoScape, Alpinion, BK Medical, Olympus and Butterfly — 915 recorded model designations in total. Every manufacturer has its own page with a searchable model list.
Next stepAll manufacturers and model designations→My model is not on the list. Do you repair it anyway?
Very probably yes. The list is a snapshot, not a limit — series get extended, renamed and designated differently by region. Send us the designation from the type plate and we will tell you whether we work on it.
Next stepSend an enquiry — answer within one working day→Where do I find the exact model designation of my probe?
On the handle or the connector, usually embossed or on a label, often together with a serial number. Take down the full designation with all suffixes — C1-5-D is not the same as C1-5.
Next stepClassify your case in the repair advisor→Are you affiliated with GE, Philips or Siemens?
No. We are manufacturer-independent and not affiliated with any equipment manufacturer. We name brands solely to identify the probes we work on.
Next stepAll manufacturers and model designations→Do I lose the manufacturer's warranty by having you repair it?
An existing manufacturer warranty or a running service contract can be affected by third-party intervention. Check your documents before placing the order. For probes outside warranty — the normal case in practice — the question does not arise.
Next stepSend an enquiry — answer within one working day→Do you use original parts?
We use components that are suitable and tested for the probe in question. What exactly was fitted is stated in the test report — ask beforehand if in doubt.
Next stepSend an enquiry — answer within one working day→Do you also sell probes, or only repairs?
This part of the business is repair. If you need a replacement or want to hand in a probe, talk to us and we will tell you what is possible.
Next stepSend an enquiry — answer within one working day→Can you work on a probe from a manufacturer whose system we no longer run?
Yes, and that is exactly where repair is often the only option: replacement probes for end-of-life systems are barely available. As long as we know the design, something can almost always be done.
Next stepAll manufacturers and model designations→How do I send the probe in?
You fill in the enquiry form — fastest through the repair advisor, which prepares it for you. On request we organise collection across Europe. Enclose the printed form with the shipment and your case goes straight to the technicians without queries.
Next stepAll manufacturers and model designations→Hygiene and safety
Disinfection before shipping, the hygiene declaration and electrical safety after the repair.
Do I have to disinfect the probe before shipping?
Yes. Every incoming probe must be reprocessed and accompanied by a hygiene declaration. That protects our technicians and is the condition for us to be allowed to accept the probe.
Next stepEnquiry protocol as a PDF→What is the hygiene declaration and where do I get it?
A short written confirmation from your institution that the probe was disinfected according to your procedures. The enquiry protocol contains a field for it — or enclose your own form.
Next stepEnquiry protocol as a PDF→How is the probe cleaned at your workshop?
The probe is reprocessed before and after every intervention, before it goes into measurement or into shipping. The test report comes back with the probe and documents the process.
Next stepSend an enquiry — answer within one working day→Is the probe electrically safe after the repair?
Every probe is only returned after the final inspection, which includes the electrical test and the visual or leak-tightness test. The results are in the test report.
Next stepSend an enquiry — answer within one working day→What happens to a probe that is no longer safe to use?
Then we say so plainly and do not half-repair it. A probe that fails the final inspection does not go back as repaired — you receive the findings and decide whether to dispose of it or have it returned.
Next stepSend an enquiry — answer within one working day→Can we use a repaired probe on patients without further steps?
Once it has passed the final inspection and comes with the test report, the probe is intended for use. Entering it into your own device documentation and any in-house release remain your task as the operator.
Next stepSend an enquiry — answer within one working day→What happens if the repair is not worthwhile?
Then we tell you. You receive the assessment with the quotation and decide. If you decline, we send the probe back. We repair nothing without your approval, and no costs arise that you did not expect.
Next stepSend an enquiry — answer within one working day→Law, documentation, warranty
MPBetreibV, test report, warranty periods, liability and data protection.
Does a repair meet the requirements of the German MPBetreibV?
In Germany and Austria the MPBetreibV requires medical devices to be operated in a safe and functional condition and maintenance to be documented (§ 5, § 7). That is exactly what you receive with every repair: a test report. Comparable rules apply in most EU countries.
Next stepSend an enquiry — answer within one working day→What is in the test report?
The condition on arrival, the measured results, the work carried out and the result of the final inspection. It is the document you take into your device documentation.
Next stepSend an enquiry — answer within one working day→Do we get the test report digitally?
Yes, as a PDF in addition on request. Note it on the enquiry protocol and it goes to the address given there.
Next stepEnquiry protocol as a PDF→How long is the warranty on the repair?
On full overhauls including crystal replacement we give twelve months of warranty. For smaller levels we state the warranty in the quotation, related to the work carried out.
Next stepThe repair levels in detail→What if the same fault reappears after the repair?
Get in touch with the reference number — then we can see from the report what was done. If the same fault reappears on the same assembly within the warranty, that is a warranty case.
Next stepSend an enquiry — answer within one working day→Is a repair a contract for work or a contract of sale?
A repair is a contract for work: what is owed is the agreed repair result, not the delivery of goods. The details are in our terms for the order.
Next stepSend an enquiry — answer within one working day→Who is liable for the probe while it is with you?
For the period in our workshop we assume the duty of care for the device handed over. For particularly valuable probes, agree the insured value with us beforehand so that transport is covered appropriately as well.
Next stepSend an enquiry — answer within one working day→How is data protection handled for the enquiry?
The enquiry form collects only what is needed to process the case — contact details, device, fault. We deliberately do not publish an e-mail address; contact runs through the form so that enquiries do not end up in spam.
Next stepSend an enquiry — answer within one working day→Do I get a test report?
Yes, with every repair. It states the fault pattern, the work carried out, the parts fitted and the final test — ready to file with your device documentation. Operators are obliged to document maintenance; the report is designed for that.
Next stepSend an enquiry — answer within one working day→Is there a warranty on the repair?
On full overhauls including crystal replacement we give twelve months warranty. On smaller repair levels the warranty follows the scope of the work; what applies is stated in the quotation.
Next stepSend an enquiry — answer within one working day→Loan probe and downtime
How to bridge the downtime and why precautionary testing is the cheapest moment.
Do you provide a loan probe?
Subject to availability, yes. Tick the requirement on the enquiry protocol and give the model or the connector type — then we will tell you what we can provide for your system.
Next stepEnquiry protocol as a PDF→Does the loan probe cost anything?
We settle that case by case together with the quotation, depending on model and duration. Ask about it before the probe comes to us.
Next stepSend an enquiry — answer within one working day→How do I bridge the downtime without a loan probe?
Three things help in practice: register for express handling, move examinations to a second system, and do not wait for total failure before sending the probe — an emerging fault can be planned around, a total failure cannot.
Next stepClassify your case in the repair advisor→Can we have a probe tested as a precaution?
Yes, and it is the cheapest moment there is. A measurement shows weak elements and incipient cable damage before they cause a failure in the middle of a clinic session.
Next stepClassify your case in the repair advisor→Do you also repair endoscopes and medical batteries?
Yes. Flexible endoscopes — gastroscopes, colonoscopes, bronchoscopes and duodenoscopes — and ultrasound endoscopes for endosonography are repaired just like medical battery packs of all technologies including lithium and lithium-ion. Battery development sits inside the group with AccuPower.
Next stepSend an enquiry — answer within one working day→Company and sustainability
Who repairs, where the work is done and what a repair saves ecologically.
Who is behind this repair business?
MOONS in Graz, which grew out of AccuPower, a manufacturer of energy and electronics solutions for medical devices since 1998. Registered as a medical device operator under GISA 33592718.
Next stepSend an enquiry — answer within one working day→Where is the repair carried out?
In our own workshop in Graz, Austria. Nothing is passed on to third parties in the background — the probe is opened, measured and tested there.
Next stepSend an enquiry — answer within one working day→What does a repair save ecologically?
Around 80 kilograms of CO₂ compared with producing a new probe, plus the raw materials and the electronics that never have to be made in the first place. Every probe that is not replaced is the most effective part of the calculation.
Next stepThe repair levels in detail→What is the fastest way to reach you?
Through the enquiry form on this site, ideally with a completed enquiry protocol — then the answer within one working day is usually already an assessment. By phone you can reach us on +43 316 22 81 10.
Next stepSend an enquiry — answer within one working day→Does the probe have to be disinfected before I send it?
Yes. The probe is handed over cleaned and disinfected, which you confirm in the enquiry form with the hygiene declaration. It comes back from us reprocessed and disinfected, documented to ISO standards on request.
Next stepSend an enquiry — answer within one working day→Nothing here on that yet. Describe the case in the enquiry form — we answer within one working day.
Request a repair →Send in a probe with this fault
Five questions in the repair advisor, and at the end you get a completed enquiry form with an initial assessment.
Request a repair