Fluid ingress.
Gel, disinfectant or body fluid enters through a leak — at the strain relief, at the lens, on TEE probes through the bending rubber. Inside, the fluid goes on corroding leads and electronics even while the probe still appears to work.
How to spot it
Condensation behind the lens, moisture in the connector, salty residues, an unusual smell. On TEE probes the leak test shows ingress reliably. Any suspicion of fluid ingress means: take the probe out of service immediately — it is an electrical risk for patient and user.
What we do
We open the probe, determine the extent of the corrosion and say honestly whether a repair is still worthwhile. Where it is, affected leads, seals and electronics are renewed, after which leakage current and tightness are tested.
Usual repair level
Every repair is individual and quoted on request — depending on the probe model and its condition. After the analysis you receive a transparent quotation. Additional services: disinfection to ISO standards, collection across Europe, 24h express service.
Probe types affected
- Standard & matrix probes
- Wireless probes
- TEE & matrix TEE probes
- Flexible endoscopes
- Ultrasound endoscopes